DPSS DMP LH series  has quite a few advantages over the Millennia Prime:
Laser Specifications
• Better power stability. The DPSS-DMP-LH10 has much better power stability than 
the Millennia: 0.5% RMS for DPSS-DMP-LH10 versus 2.0% RMS for Millennia.
• Lower optical noise. Even though the noise specifications are, on paper, quite 
similar: 0.03% RMS for DPSS-DMP-LH10 versus 0.04% RMS for Millennia, we have 
heard from a number of customers who have tested both lasers that, in reality, 
the DPSS-DMP-LH10 noise is much lower. It is also quite well known that the 
Millennia has a high frequency spatial instability.
· Better spectral purity. The Millennia has no specification on the amount of 
1064nm light contained in the 532nm output beam. Typically it can be hundreds of 
milliwatts or more. Most times this 1064nm light ends up being scattered inside 
the Ti:Sapphire laser or around the laboratory causing potential eye hazards. 
The DPSS-DMP-LH10 has a spectral purity specification of > 99.9%. This means 
that even in the 10W, 532nm output there is at most only 10mW of 1064nm light.
more
Laser Head: Physical Specifications & Features
• Vents on laser head. There are heat vents on the Millennia laser head to allow 
excess heat to escape from the intra-cavity frequency-doubler assembly. This 
excess heat must then be removed by the customer. The DPSS-DMP-LH10 laser head 
is hermetically sealed with no vents allowing the laser head to be completely 
integrated inside an OEM product.
• Laser head dimensions. The volume of the DPSS-DMP-LH10 laser head is less than 
50% of the Millennia laser head: 144 cubic inches for the DPSS-DMP-LH10 versus 
304 cubic inches for the Millennia.
• Laser head weight. The weight of the DPSS-DMP-LH10 laser head is ~ 30% that of 
the Millennia: 7.0 lbs for the DPSS-DMP-LH10 versus 22.0 lbs for the Millennia.
more
Power Supply-Cooler: Physical Specifications & Features
· Pump diode(s). All DPSS-DMP-LH models (5W, 6W, 8W and 10W) use one pump diode 
module that is housed in the power supply. The armored delivery fiber is easily 
disconnected at the diode module for field servicing. The pump diode modules for 
the Millennia, also housed in the power supply, have permanently-attached fiber 
assemblies that are disconnected at the Millennia laser head for field 
servicing. During this service procedure the fiber assembly must be removed from 
the long umbilical that connects the Millennia laser head, power supply and 
water chiller.
· Integrated cooler (chiller). The DPSS-DMP-LH10 has a purpose-built, 
integrated, closed-loop chiller. The Millennia comes with a large, separate, 
stand-alone chiller. The physical dimensions for the Millennia chiller are not 
given anywhere on the Millennia datasheet.
· Power consumption. The power consumption figures for the Millennia chiller are 
not included in those given on the Millennia datasheet. The power consumption 
for the complete DPSS-DMP-LH10 system, including the integrated chiller, are 
less than 500 Watts. 
more
Cost -
request a quote
• Initial purchase price. The DPSS-DMP-LH10 price is much less than the 
Millennia.
• Servicing the diode pack. The diode pack for the DPSS-DMP-LH10, contained in 
the power supply, is inexpensive and can be replaced by the customer. The diode 
packs for the Millennia are expensive and can only be replaced by a 
Spectra-Physics service engineer trained on the Millennia.